Replacing the purge kit actuator

Important

Before replacing the actuator, unplug the purge actuator kit from the Asycube, remove the platform and remove the actuator kit.

Step 1:

Remove the protective cover (A):

4 screws Tx10. torx

../../_images/replacing_kit_step1.png

Step 2:

Disconnect the actuator by pulling on the connector, and not on the cable itself (B).

../../_images/replacing_kit_step2.png

Step 3:

Unscrew the actuator block:

4 screws Tx10 torx

../../_images/replacing_kit_step3.png

Step 4:

Unscrew and open the support (1):

4 screws X0. cruciform

Unscrew the actuator (2):

1 screws Tx8. torx

Swap out the actuator (C).

../../_images/replacing_kit_step4.png

Step 5:

Screw back the actuator in the support (1):

1 screws Tx8 - 0.8Nm. torx

Close and screw support (2):

4 screws X0 - 0.6Nm. cruciform

../../_images/replacing_kit_step5.png

Step 6:

Screw back the actuator block on the main plate:

4 screws Tx10 - 1Nm torx

../../_images/replacing_kit_step6.png

Step 7:

Plug in the actuator (D).

../../_images/replacing_kit_step7.png

Step 8:

Screw the protective cover (E) back into position:

4 screws Tx10 - 0.6Nm. torx

../../_images/replacing_kit_step8.png